14 Settings
UMFLUXUS_F7V4-6-2EN, 2017-10-01
133
14.4
Setting the Contrast
It is possible to reset the display to medium contrast. Press key C. Enter HotCode
555000
.
14.5
Instrument Information
Select
Special Funct.\SYSTEM settings\Miscellaneous
to set the contrast of
the display of the transmitter. Press ENTER.
The contrast of the display is adjusted with the following keys:
increases the contrast
reduces the contrast
= min. contrast
= medium contrast
= max. contrast
Note!
After an initialization of the transmitter, the display is displayed to medium contrast.
Select
Special Funct.\Instrum. Inform.
to display information about the transmit-
ter. Press ENTER.
The type and the serial number of the transmitter will be displayed in the upper line.
The max. available data logger memory will be displayed in the lower line (here: 18 327 ad-
ditional measured values can be stored). For further information on the data logger see
section 12.1.6.
Press ENTER.
The type and the serial number of the transmitter will be displayed in the upper line.
The firmware version of the transmitter with date is displayed in the lower line.
Press ENTER.
SYSTEM settings
↕
Miscellaneous
SETUP DISPLAY
←
CONTRAST
→
6
4
2
5
8
Special Funct.
↕
Instrum. Inform.
F70X -XXXXXXXX
Free:
18327
F70X -XXXXXXXX
V x.xx
dd.mm.yy