A Menu Structure
UMFLUXUS_F7V4-6-2EN, 2017-10-01
169
input of the lowest/highest measured value to be expected for the current output
The values are assigned to the lower/upper limit of the output range.
These displays will only be indicated if
Current Loop
is activated.
input of the error value delay, i.e. of the time interval after which the value en-
tered for the error output will be transmitted to the output if no valid measured
values are available
This display will only be indicated if
Special Funct.\SYSTEM settings\
Dialogs/Menus\Error-val. delay
=
EDIT
has been selected.
Pulse Output
Activation of a Pulse Output
This display will only be indicated if a pulse output has been installed in
Spe-
cial Funct.\SYSTEM settings\Dialogs/Menus\Proc. outputs
.
input of the pulse value (value of the totalizer at which a pulse will be emitted)
This display will only be indicated if
Pulse Output
is activated.
input of the pulse width
range: 1 or 80...1000 ms
This display will only be indicated if
Pulse Output
is activated.
Alarm Output
activation of an alarm output
This display will only be indicated if an alarm output has been installed in
Spe-
cial Funct.\SYSTEM settings\Proc. outputs
.
Selection of the switching condition (
func
), the holding behavior (
typ
) and the
switching function (
mode
) of the alarm output.
This display will only be indicated if
Alarm Output
is activated.
selection of the physical quantity to be monitored
This display will only be indicated for R1 if
Alarm Output
is activated.
input of the upper limit of the physical quantity to be monitored
This display will only be indicated if
Alarm Output
has been activated and
MAX
has been selected as the switching condition.
input of the lower limit of the physical quantity to be monitored
This display will only be indicated if
Alarm Output
has been activated and
MIN
has been selected as the switching condition.
input of the limit for the totalizer of the physical quantity to be monitored
This display will only be indicated if
Alarm Output
has been activated and
QUANT.
has been selected as the switching condition.
input of the hysteresis for the lower or upper limit
This display will only be indicated if
Alarm Output
has been activated and
MIN
or
MAX
has been selected as the switching condition.
Program Branch
Special Funct.
main menu: selection of the program branch
Special Funct.
INIT-
resistant
Zero-Scale Val.
0.00
m3/h
Full-Scale Val.
300.00
m3/h
Error-val. delay
10
s
Pulse Output
B1: no
>YES<
Pulse Value
0.01
m3
Pulse Width
100
ms
Alarm Output
no
>YES<
R1=FUNC<typ mode
Function:
MAX
R1 Input:
↕
Volume flow
High Limit:
-10.00
m3/h
Low Limit:
-10.00
m3/h
Quantity Limit:
1.00
m3
R1 Hysterese:
1.00
m3/h
par mea opt >SF<
Special Funct.