A Menu Structure
168
UMFLUXUS_F7V4-6-2EN, 2017-10-01
input of the number of sound paths
display of the transducer distance to be adjusted between the inner edges of the
transducers
Program Branch
Output Options
main menu: selection of the program branch
Output Options
selection of the channel whose output options are to be defined
selection of the physical quantity
selection of the unit of measurement for the physical quantity
activation of a temperature input
This display will only be indicated if the temperature input T1 has been assigned
to the channel in
Special Funct.\SYSTEM settings\Proc. inputs\
Link temperature
.
activation of a current input for an external temperature measurement
This display will only be indicated if the input I1 has been assigned to the chan-
nel in
Special Funct.\SYSTEM settings\Proc. inputs\Link other
inp.
.
input of the duration over which a floating average of the measured values has
to be determined
range: 1...600 s
activation of the data logger
activation of the measured value transmission to a PC or a printer via the serial
interface
selection of the storage rate for storing measured values in the data logger
This display will only be indicated if
Output Options\Store Meas.Data
and/or
Serial Output
are activated.
Input of the storage rate if
Storage Rate
=
EXTRA
has been selected
range: 1...43 200 s (= 12 h)
Current Loop
activation of a current output
This display will only be indicated if the current output has been installed in
Special Funct.\SYSTEM settings\Proc. outputs
.
selection whether the sign of the measured values is to be considered for the
output
This display will only be indicated if
Current Loop
is activated.
INIT-
resistant
A: Sound Path
2
NUM
Transd. Distance
A:54 mm Reflex
par mea >OPT< sf
Output Options
Output Options
↕
for Channel
A:
Physic. Quant.
↕
Volume flow
Volume in:
↕
m3/h
Temperature
T1
no
>YES<
Input
I1
no
>YES<
Damping
10
s
Store Meas.Data
no
>YES<
Serial Output
no
>YES<
Storage Rate
↕
Once per 10 sec.
Storage Rate
1
s
Current Loop
I1: no
>YES<
Meas.Values
>ABSOLUT<
sign