18 Outputs
UMFLUXUS_F7V4-6-2EN, 2017-10-01
153
Tab. 18.3: Examples for the error output
list item for the error output
output signal
error output = 2 mA
Select a list item for the error output. Press ENTER.
If
Other value
has been selected, enter an error value. It has to be within the limits of the
output.
Press ENTER.
Note!
The settings will be stored at the end of the dialog.
The terminals for the connection of the output are displayed (here:
P1+
and
P1-
for the ac-
tive current loop).
Press ENTER.
Error-value
↕
Minimum (4.0mA)
I [ m A ]
2 0
4
t
d
Error-value
↕
Hold last value
I [ m A ]
2 0
4
Error-value
↕
Maximum (20.0mA)
I [ m A ]
2 0
4
t
d
Error-value
↕
Other value...
I [ m A ]
2 0
4
t
d
Error-value
↕
Minimum (4.0mA)
Error-value
3.5
mA
I1 active loop
Terminal:P1+,P1-