6 Installation
44
UMFLUXUS_F7V4-6-2EN, 2017-10-01
• Put the rail on the tension strap clamps (see Fig. 6.51).
• Fix the rail to the tension strap clamps with the screws.
• Tighten the screws.
6.4.4.3
Installation of the transducers in Variofix C
• Put coupling foil (or some coupling compound for a short-term installation) on the contact surface of the transducers. The
coupling foil can be fixed to the contact surface with a small amount of the coupling compound.
• Position the transducers on the rail in such way that the engravings on the transducers form an arrow. The transducer
cables show in opposite directions (see Fig. 6.52).
• Adjust the transducer distance displayed by the transmitter (see section 9.5 and Fig. 6.52).
• Slide the spring clips on the transducers (see Fig. 6.53).
• Fix the transducers by tightening the tensioning screws slightly. The end of the screw has to be placed above the hole in
the transducer (see Fig. 6.52).
• Correct the transducer distance, if necessary (see section 9.6.1 and section 9.6.2).
• Tighten the tensioning screw.
• Fix the spacing element on the rail to mark the transducer position (see Fig. 6.52).
• Use a cable tie to fix the transducer cables in order to protect them from mechanical strain (see Fig. 6.53).
• Put the cover on the rail (see Fig. 6.54).
• Tighten the screws on both sides of the cover.
Fig. 6.50: Ratchet clasp with tension strap
Note!
To release the screw and the tension strap press the lever down (see Fig. 6.50).
Fig. 6.51: Rail on the pipe
Note!
If coupling foil is used: If the signal is not sufficient for the measurement, use the coupling compound
instead of the coupling foil.
lever
sense of rotation
cut
screw of the clasp
screws