18 Outputs
154
UMFLUXUS_F7V4-6-2EN, 2017-10-01
18.1.3
Function Test
The function of the installed output can now be tested. Connect a multimeter with the installed output.
Test of the Analog Outputs
Test of the binary outputs
18.2
Error Value Delay
The error value delay is the time interval after which the error value will be transmitted to the output in case no valid mea-
sured values are available. The error value delay can be entered in the program branch
Output Options
if this menu
item has previously been activated in the program branch
Special Funct.
. If the error value delay is not entered, the
damping factor will be used.
18.3
Activation of an Analog Output
The current output is tested in the display. Enter a test value. It has to be within the output
range. Press ENTER.
If the multimeter displays the entered value, the output functions correctly.
Select Sie
yes
to repeat the test,
no
to return to
SYSTEM settings
. Press ENTER.
Select
Reed-Relay OFF
or
Open collect OFF
in the scroll list
Output Test
to test
the de-energized state of the output. Press ENTER. Measure the resistance at the output.
The value has to be high ohmic.
Select
yes
. Press ENTER.
Select
Reed-Relay ON
or
Open collect. ON
in the scroll list
Output Test
to test the
energized state of the output. Press ENTER. Measure the resistance at the output. The val-
ue has to be low ohmic.
Select
yes
to repeat the test,
no
to return to
SYSTEM settings
. Press ENTER.
Select
Special Funct.\SYSTEM settings\Dialogs/Menus\Error-val. de-
lay.
Select
Damping
if the damping factor is to be used as the error value delay. Select
Edit
to
activate the menu item Error-val. delay in the program branch
Output Options
.
From now on, the error value delay can be entered in the program branch
Output Op-
tions
.
Note!
An output can only be activated in the program branch
Output Options
if it has previously been in-
stalled.
In the program branch
Output Options
, select the channel for which an output is to be
activated. Press ENTER.
This display will not be indicated, if the transmitter has only one measuring channel.
Press ENTER until
Current Loop
is displayed. Select
yes
to activate the output. Press
ENTER.
I1:Output Test
4
mA
I1= 4.0 mA
Again? no
>YES<
B1:Output Test
↕
Reed-Relay OFF
B1=OFF
AGAIN? no
>YES<
B1:Output Test
↕
Reed-Relay ON
B1=ON
AGAIN? no
>YES<
Error-val. delay
>DAMPING<
edit
Error-val. delay
10
s
Output Options
↕
for Channel
A:
Current Loop
I1: no
>YES<