Enter the half PD and the frame height.
Change the distance between the PRP (prismatic reference point) and the centering cross
if necessary.
This value is set by default to 6 mm for Essilor mid-range lenses. This can vary depending on the
lens manufacturer.
The target moves.
Make sure that the standard tripod is inside the centering chamber.
•
Use the high-base tripod for a high-base lens.
•
Place the recut accessory in the center of the standard tripod for a small lens.
Position the lens on the tripod.
Move the lens slowly until the manufacturer markings are superimposed on the target.
>
3
4
5
6
7
U
SER
M
ANUAL
> III. C
ENTERING
A
LENS
65
Pro-B 300 > v1.0 - 03.19