Waterpilot FMX21
Description of device parameters
Hauser
119
14.16 Expert → Application
Electr. Delta P
Write permission
Operators/Service engineers/Expert
Description
For switching the Electr. Delta P application off or on with an external or constant value.
Options
• Off
• External value
• Constant
Factory setting
Off
Fixed ext. value
Write permission
Operators/Service engineers/Expert
Description
Use this function to enter the constant value. The value refers to "HART input unit"
Factory setting
0.0
Auto dens. corr.
Write permission
Operators/Service engineers/Expert
Description
For switching the auto dens. corr. application off or on with an external or internal
temperature value.
Before performing a calibration (dry or wet), auto-density compensation must be switched
on if this function is to be used. As soon as "Auto dens. corr." is switched on, the field for
entering the "Process density" and "Adjust density" is disabled.
The calibration density remains the last value until it is overwritten by a calibration. The
process density remains the last value until it is overwritten when the system recalculates
the value.
Automatic density compensation is performed for the 0 to 70 °C (32 to 158 °F)
temperature range. The density values for water are used for this density compensation.
Prerequisite
Level mode
Options
• Off
• Sensor temperature
• External value (only if Off or Constant is selected for Electr. Delta P)
Factory setting
• Off
• On (if the option "IC" was selected in the "Service" order code when ordering)