14.8 Expert → Measurement → Linearization
Lin. mode
Write permission
Operators/Service engineers/Expert
Description
Select the linearization mode.
Options
• Linear
The level is output without being converted beforehand. "Level before lin" is output.
• Erase table
The existing linearization table is deleted.
• Manual entry (sets the table to edit mode, an alarm is output):
The value pairs of the table (X-value and Y-value) are entered manually.
• Semi-automatic entry (sets the table to edit mode, an alarm is output):
The vessel is emptied or filled in stages in this entry mode. The device automatically
records the level value (X-value). The associated volume, mass or % value is entered
manually (Y-value).
• Activate table
The table entered is activated and checked with this option. The device shows the level
after linearization.
Factory setting
Linear
Unit after lin.
Write permission
Operators/Service engineers/Expert
Description
Select volume unit, mass, height or % (unit of the Y-value).
Options
• %
• cm, dm, m, mm
• hl
• in
3
, ft
3
, m
3
,
• l
• in, ft
• kg, t
• lb
• gal
• Igal
Factory setting
%
Line-numb