Description of device parameters
Waterpilot FMX21
100
Hauser
Description
Enter a new density value for density correction. The calibration was carried out with the
medium water, for example. Now the vessel is to be used for another medium with
another density. The calibration is corrected appropriately by entering the new density
value in the "Process Density" parameter.
Input: Auto dens. corr. = Off
Display: Auto dens. corr. ≠ Off
Note
If, after completing a wet calibration, you change to dry calibration using the "Calibration
mode" parameter, the density for the "Adjust density" and "Process density" parameters
must be entered correctly before changing the calibration mode.
Factory setting
1.0
Level before lin.
Write permission
No write permissions. Parameter is read only.
Description
Displays the level value prior to linearization.