Installation
16
Proceed as follows to wall-mount the unit:
1.
Bore the holes as per Fig. 10.
2.
Push the two securing screws from the front through the securing bores (item 1).
3.
Mount the transmitter housing to the wall as illustrated.
4.
Cover the bores with the plastic caps (item 2).
3.4.4
Post mounting and panel mounting
!
Note!
You need a special mounting kit to secure the transmitter to horizontal and vertical posts or pipes
(max. Ø 70 mm (2.76")) and for panel mounting.
a0003955
Fig. 11:
Mounting kit
Panel mounting
Proceed as follows to panel-mount the transmitter:
a0003956
Fig. 12:
Panel mounting
1.
Make the necessary mounting cutout, measuring 161 x 241 mm (6.34" x 9.49").
The installation depth is 134 mm (5.28").
2.
Unscrew the top housing section (item 1).
3.
Secure the securing plates (item 3) to the transmitter housing base using the securing screws
(item 5) in accordance with Fig. 12.
4.
Secure the transmitter to the panel (item 2) using the clamping screws (item 4).
5.
Place the flat seal (see "Accessories" section) on the housing base.
6.
Screw the top housing section back on.