Wiring
29
a0003977
Fig. 28:
pH terminal on housing cover
4.
Open the top housing section of the device.
5.
On the right-hand side of the housing cover, disconnect the red cable to the pH input at both
ends (see Fig. 29).
6.
Attach the jumpers supplied as illustrated in Fig. 30.
7.
Connect the sensor cable in accordance with the ISFET assignment.
8.
In the Quick Setup, change the electrode type to "ISFET".
!
Note!
Proceed accordingly for changing from ISFET sensors to glass electrodes.
pH
Ref
PA
12
13
11
S
pH
a0003978
Fig. 29:
pH input module in the housing cover with
cable (red) for connecting glass electrodes
a0003979
Fig. 30:
pH input module in the housing cover with
jumper for connecting ISFET sensors