Troubleshooting
143
10.4
Device-specific errors
The following table supports you during diagnosis and provides information on the necessary spare
parts, where applicable.
Depending on the degree of difficulty and measuring media available, diagnosis is carried out by:
• Specialized staff of the owner-operator
• Electrical technicians of the owner-operator
• System creator/operator
• Hauser Service
Information on the exact names for spare parts and the installation of these parts is found in the
"Spare parts" section.
Assembly moves several
times without correct
feedback
Pneumatic connections mixed up
See "Pneumatics and hydraulics, CPG"
section.
• Hose 2 "Measure" at valve 5 at front
• Hose 3 "Service" at valve 5 at rear
Feedback not correctly connected
See "Pneumatics and hydraulics, CPG"
section.
• Hose 5 "Measure feedback" (term.
11/12)
• Hose 6 "Service" feedback (term. 13/14)
Feedback switch misaligned
Feedback switch under pressure
• Open: Term. 11/12-13/14 = 14 V
• Closed: Term. 11/12-13/14 = 0 V
Assembly was moved manually.
Feedback signal is thus undefined.
Position change via manual operation menu
Error
Possible cause
Tests and / or remedial action
Equipment needed, spare parts
Error
Possible cause
Tests and / or remedial action
Executed, equipment needed, spare
parts
Display dark, no light
emitting diode active
No line voltage
Check whether line voltage present.
Electrician / e.g. multimeter
Incorrect supply voltage or too low
Compare actual line voltage and
information on the nameplate.
User
(Energy supply company or multimeter data)
Connection defective
Terminal not tightened;
Insulation jammed;
Incorrect terminals used.
Electrical technician
Device fuse defective
Compare line voltage and information on
the nameplate and replace fuse.
Electrician/suitable fuse;
See "Spare parts" section.
Power unit defective
Replace power unit. Pay attention to
version.
Local diagnosis: All 6 LEDs on the M3G
module must light up.
Central module defective (if all 6 LEDs
in the M3G power unit light up)
Replace central module. Pay attention to
version.
Diagnosis by Hauser Service on
site, test module required
Ribbon cable loose or defective
Check ribbon cable, replace if necessary.
Cable on the side of the M3G module
welded in.
Display dark, light emitting
diode active
Central module defective (module:
M3Cx-x)
Replace central module M3Cx-x. Pay
attention to version.
Diagnosis by Hauser Service on
site, test module required
Display shows value but
– no change in display
and/or
– Device cannot be
operated
Instrument or module in instrument not
correctly installed
Check module connections
Perform with the aid of the device view in
the "Spare parts" section.
Operating system in impermissible state Switch device off and on again.
Possible EMC problem: If the problem
persists, check installation or have checked
by Hauser service.
Device gets hot
Voltage incorrect/too high
Compare line voltage and information on
the nameplate.
User, electrical technician
Power unit defective
Replace the power unit.
All 6 red LEDs on the M3G module must
light up.