Wiring
30
4.1.7
Digital sensors with Memosens technology
Measuring cables
You require the Memosens CYK10 data cable to connect digital sensors:
Preparing cables
a0003976
Fig. 31:
Outer screen connection with metal cable gland
1.
Slide the cable gland and the clamping ring over the cable.
2.
Remove the inner insulation.
3.
Remove the outer shield from the cable and fold it back over the clamping ring.
4.
Guide the sensor cable through the cable opening of the device and screw the gland closed.
Shield contacting takes place automatically here.
Sensor type
Cable
Extension
Digital sensors with temperature sensor
CYK10
RM junction box + CYK81 cable
1
3
4
2