Commissioning
112
Simulation
Simulation of current outputs
Activation and subsequent
entry
• Simulation
– On
– Off
• Output 1
0.0 ... 22.0 mA
• Output 2
0.0 ... 22.0 mA
Factory setting
Simulation: Off
Output 1: 0.0 mA
Output 2: 0.0 mA
Adjust the simulation of the current outputs.
Simulation off:
The frozen values of the last
measurement are used for simulation.
Simulation on:
The current values for the
outputs can be altered for the simulation
(output 1, output 2).
Simulation of measured value,
temperature
Activation and subsequent
entry
• Simulation
– On
– Off
• pH/mV 1
-2.00 ... 16.00 pH
• Temperature
-50 ... +150
°
C
Factory setting
Simulation: Off
pH/mV 1: 7.00 pH
Temperature: 25.0
°
C
Adjust the simulation of measured values and
the temperature.
Simulation off:
The frozen values of the last
measurement are used for simulation.
Simulation on:
The values can be changed for
the simulation.
Simulation of relays
Activation and subsequent
entry
• Simulation
– On
– Off
• Alarm relay
– On
– Off
• Relay 1
– On
– Off
• Relay 2
– On
– Off
Factory setting
Simulation: Off
Alarm relay: Off
Relay 1/2: Off
Adjust the simulation of the relays.
Simulation off:
The last states are frozen and
used for simulation.
Simulation on:
The relays can be opened (on)
or closed (off).
!
Note!
If you return to the measurement mode with
the simulation switched on, "Simul" and "Hold"
flash in the display.
Instrument check
Select test
Options
• Display
• Keypad
• RAM
• EEPROM
• Flash
You can check the function of the transmitter
with the instrument check.
Display:
All the fields of the display are
addressed alternately. In this way, any defective
cells are visible.
Key pad:
The keys all have to be pressed one
after the other. If the system is functioning
perfectly, the appropriate symbols appear in the
display.
RAM:
"RAM o.k." message if operating
correctly without error.
EEPROM:
"EEPROM o.k." message if
operating correctly without error.
Flash
(memory): "Flash o.k." message if
operating correctly without error.
Function
Options
Info