Installation
15
3.4.3
Wall mounting
"
Caution!
• Make sure the maximum permitted ambient temperature range of -20 ... +60
°
C (-4 ... 140
°
F)
is observed. Avoid direct sunlight.
• Always mount the wall-mount housing in such a way that the cable entries point downwards.
Control unit
a0004861
Fig. 9:
Dimensions for wall mounting with wall securing kit (part of scope of delivery)
Proceed as follows to wall-mount the unit:
1.
Please note that the maximum suction height for buffer and cleaner is 2.5 m (8.2 ft.) when
using the standard multihoses supplied. Bore the holes as per the graphic above.
2.
Screw the elements of the wall securing set supplied to the rear wall of the housing.
3.
Secure the housing to the wall without any inclination.
Transmitter
a0003953
Fig. 10:
Dimensions for wall mounting, securing screw: Ø 6 mm (0.24"), wall plug: Ø 8 mm (0.31")
1
Securing bores
2
Plastic caps
154
(6.06)
100 (3.94)
247
(9.72)
142 (5.59)
167 (5.57)
Ø
8
(0.31)
2
1
mm (inch)