Commissioning
76
Redox operating mode
To call the menu, select:
"
Ñ
> Specialist > Set up 1 > Calibration"
Autocal. Topcal
Calibration parameters
Options
• Fixed buffer
• Auto. buffer recognition
Factory setting
Fixed buffer
Specify the type of calibration for automatic
calibration.
Buffer table:
This function can be selected if
the same buffer values are constantly used.
Auto. Buffer recognition:
The Mycom S
transmitter automatically recognizes the buffer
values used.
!
Note!
Automatic buffer recognition only works if glass
electrodes are connected to both measuring
circuits. If you are using an ISFET sensor, please
calibrate with another calibration function.
Buffer type
Options
• DIN 19267
• E+H
• NBS / DIN 19266
• Merck+Riedel
• Special buffer
Factory setting
E+H
Select a buffer type where the pH values
specified are fixed.
Spec. buffer table:
The tables for special
buffer, which are to be defined in the "Special
buffer" option, are used.
!
Note!
The buffer tables for the buffer types on offer
are provided in the appendix.
Buffer 1
Options
• Buffer 2.0
• Buffer 4.01
• Buffer 6.98
• Buffer 9.18
• Buffer 10.90
(option depends on the buffer
type)
Factory setting
Buffer 6.98
Select the pH value for buffer 1 of the two-point
calibration.
Buffer 2
Options
• Buffer 2.0
• Buffer 4.01
• Buffer 9.18
• Buffer 10.90
(option depends on the buffer
type)
Factory setting
Buffer 4.01
Select the pH value for buffer 2 of the two-point
calibration.
Function
Options
Info
Calibration
Options
• Offset
• Onsite calibration type
• Cal settings
• Calibration timer
• Autocal. Topcal
Factory setting
Offset
Select the initial settings for the calibration.
Offset:
Enter a fixed value by which the mV
value is shifted.
Manual calibration:
Make initial settings for
the function of the CAL key.
Cal settings:
General calibration settings
Calibration timer:
Timer for calibration
intervals
Autocal. Topcal:
Make initial settings for the
Topcal S method of calibration.
Function
Options
Info