Wiring
40
4.2
Post-connection check
After the electrical connection, carry out the following checks:
Device condition and specifications
Notes
Are the transmitter and cable damaged externally?
Visual inspection
Electrical connection
Notes
Does the supply voltage match the specifications on the nameplate?
100 ... 230 V wide area
24 V AC / DC
Do the cables used fulfill the required specifications?
Use a genuine Hauser cable for
connecting the sensor, see "Accessories"
section.
Are the mounted cables strain-relieved?
Is the cable type route completely isolated?
Along the whole cable length, run the
power supply and signal line cables
separately to avoid any mutual influence.
Separate cable channels are best.
No loops and crossovers in the cable run?
Are the signal cables correctly connected as per the wiring diagram?
Are all the screw terminals tightened?
Are all the cable entries installed, tightened and sealed?
Cable run with "water trap"?
"Water trap": Cable circuit hanging down
so that water can drip off.
Are the PE distributor rails grounded (if present)?
Grounding takes place at the point of
installation
Are all the housing covers installed and firmly tightened?
Check seals for damage.