Commissioning
59
Contrast
Contrast setting where applicable
You can increase or reduce the contrast of the
display with the arrow keys
V
and
W
.
Date + time
Weekday
Day
Month
Year
Time
Enter the complete date and time here.
These data are used for the logbooks and the
automatic cleaning system.
Mode
Options
• pH
• Redox mV
• Redox %
Factory setting
pH
!
Note!
• If you change the operating mode, all the
user settings are automatically reset. If you
want to save your settings, use a DAT
module.
• The Redox/ORP:% mode is not available for
redox sensors with Memosens technology
where the SW version is
≤
2.01.00.
Electrode type CH1
Options
• Glass el. 7.0
• Glass el. 4.6
• Antimony
• ISFET
Factory setting
Glass 7.0
Specify the sensor that is used (only pH).
!
Note!
• The type of electrode does not have to be
selected for digital sensors.
• The temperature sensor defaults to Pt 1000
when you change from a glass or antimony
electrode to ISFET. Pt 100 is selected when
the situation is the reverse.
• In the glass/ISFET version, Mycom S is
supplied for measuring with glass electrodes
as standard.
Connection type
Options
• Solution ground
• No solution ground
Factory setting
Solution ground
Specify whether you are measuring
symmetrically (=with PML) or unsymmetrically
(=without PML).
!
Note!
• This selection is not displayed for digital
sensors with Memosens technology. A
symmetrically high-resistance connection is
not required for digital data transmission.
• More information on symmetrical and
unsymmetrical measurement is provided on
the CD-ROM.
Temperature unit
Options
•
°
C
•
°
F
Factory setting
°
C
Temperature compensation,
temp.sensor input 1
Options
• ATC C1
• MTC
• MTC+Temp
Factory setting
ATC C1
Only for pH measurement:
ATC:
Automatic temperature compensation
using temperature sensor
MTC:
Temperature compensation with manual
input
MTC+temp:
Temperature compensation with
temperature entered manually. The
temperature measured with a temperature
sensor is displayed, however
Function
Options
Info
Hold
pH
7. 00
Param
Edit
[+-]
[E]
Contrast
Next
Hold
30
04
01
12:00
pH
7. 00
Param
[¯ ]
[E]
Date+time
Weekday :
Day
:
Month
:
Year
:
Next
Time
:
Mo
Select
pH
7. 00
Hold
Param
pH
Edit
[¯ ]
[E]
Sensor input
Redox/ORP mV
Redox/ORP %
Next
Hold
pH
7. 00
Param
Edit
Glas El. 7. 0
Glas El. 4.6
Antimon
ISFET
[¯]
[E]
pH electr.typeK1
Next
Hold
Edit
pH
7. 00
Param
[¯]
[E]
Sensor ground
solution ground
no solution ground
Next
Hold
pH
7. 00
Param
Edit
°C
°F
[¯]
[E]
Temp. unit
Next
Hold
Next [E]
Param
Temp.comp. C2
ATC C1
MTC
MTC+Temp
Edit [¯]
pH 7.00