Troubleshooting
140
10.3
Process-specific errors
Use the following table to locate and eliminate any errors occurring.
Maintenance
0
E176
Not used
Maintenance
1
E177
SCS reference electrode warning 1 (digital sensor)
Maintenance
2
E178
Not used
22
Failure
3
E179
Not used
Failure
4
E180
Data error sensor 1 (digital sensor)
Maintenance
5
E181
Not used
Maintenance
6
E182
Not used
Maintenance
7
E183
Not used
Byte
NAMUR
Bit
Error no.
Short error description
Error
Possible cause
Tests and / or remedial action
Equipment needed, spare parts
Device cannot be operated,
display value 9999
Operation is locked via key pad ("CAL"
+ "DIAG" pressed simultaneously =
locked)
Press "MEAS" and "PARAM" simultaneously
to unlock.
Measuring chain zero-point
cannot be set
Reference system poisoned
Test with a new sensor
pH/redox sensor
Diaphragm blocked
Clean or grind diaphragm
HCl 3 %, file (only for ceramic diaphragm,
only file in one direction)
Measuring line broken
Short-circuit pH input at device
⇒
Display
pH 7
Asymmetrical sensor voltage too great
Clean diaphragm or test with another
sensor
HCl 3 %, file (only for ceramic diaphragm,
only file in one direction)
Potential matching (PA/PM)
transmitter
⇔
Incorrect medium
Unsymm.: No PM or PM at PE
Symm.: PM connection mandatory
See "Connecting analog sensors" section
No calibration possible as
sensor adjustment time too
long
With ISFET sensor: Film of moisture on
measuring surface removed by drying or
blowing out with compressed air
Ensure film of moisture or buffer dwell time
> 6 min.
No or creeping display
change
Sensor fouled
Clean sensor
See "Cleaning pH/redox electrodes" section.
Sensor old
Replace sensor
New sensor
Sensor defective (reference lead)
Replace sensor
New sensor
Problem with diaphragm or missing
electrolyte
Check KCl supply (0.8 bar above medium
pressure)
KCl (CPY4-x)
Measuring chain slope
cannot be adjusted/slope
too low
Connection not at high impedance
(humidity, dirt)
Test cable, connector and junction boxes
pH simulator, insulation, see "Checking the
connecting cables and junction boxes"
section
Instrument input defective
Directly test instrument
pH simulator
Sensor old
Replace sensor
pH sensor
Measuring chain slope
cannot be adjusted/no
slope
Hairline crack in glass membrane
Replace sensor
pH sensor
Connection not at high impedance
(humidity, dirt)
Test cable, connector and junction boxes
pH simulator, insulation, see "Checking the
connecting cables and junction boxes"
section
Semiconductor layer in measuring cable
not removed
Check inner coaxial cable, remove black
layer