Commissioning
69
Temperature compensation menu
To call the menu, select:
"
Ñ
> Specialist > Set up 1 > Temperature"
Function
Options
Info
Temperature measurement
Options
• Off
• On
Factory setting
On
Only available with redox measurement
Select temperature compensation
Options
• Temp. comp. sensor
• Temp. comp. process (only for
pH operating mode)
Factory setting
Temperature
Only available with pH measurement
Select the temperature compensation
necessary.
Temperature compensation sensor:
Automatic (ATC) or manual (MTC)
temperature compensation
Temperature compensation process:
Compensation of the medium temperature
using customer-specific tables (see below)
Temperature
Temperature compensation
Options
• ATC C1
• MTC
• MTC+Temp
Factory setting
ATC C1
Only for pH measurement:
ATC:
Automatic temperature compensation
using temperature sensor
MTC:
Temperature compensation with manual
input
MTC+temp:
Temperature compensation with
temperature entered manually. The
temperature measured with a temperature
sensor is displayed, however
!
Note!
The type of temperature compensation selected
here is active during measuring operation. For
the calibration, you must configure the desired
compensation in the "Calibration" menu.
MTC temperature
-20.0 ... 150.0
°
C
Factory setting
25.0
°
C
Only available with pH and MTC or
MTC+Temp selected in the previous field.
Temperature sensor
Options
• Pt100
• Pt1000
• NTC 30K
Factory setting
Pt100
Select the temperature sensor to suit your
sensor
Temperature actual value
-5.00 ... +5.00
°
C
Factory setting
0.00
°
C
You can change/adjust the value currently
measured with the temperature sensor. The
temperature difference is stored internally as an
offset value.
Offset value
-5.0 ... +5.0
°
C
You can edit or reset the offset value resulting
from the previous field here.
Temperature compensation process
Select compensation
Options
• Select temp. table
• Edit table
• Reference temperature
Factory setting
Select temp. table
Select temp. table
You activate an existing
table.
Create table:
You create a specific table to suit
your requirements.
Reference temperature:
You enter a
temperature to which your measured values
refer.
Hold
mV -
114
Param
Edit
[¯]
[E]
Temp.meas1
Next
on
off
Hold
pH
7. 00
[E]
Select
Next
Edit[¯]
Param
Temp. comp. sensor
Temp. comp. process