Solimotion FTR20
Commissioning
30
Hauser
8.3.2
Calibration with no movement of bulk solids or minimum solid flow
The automatic calibration with no movement of bulk solids or minimum solid flow (
function
2
) is adequate for most applications. If movement is detected anyways after an automatic
calibration with no movement of bulk solids (for example, movement in the area surrounding
the measuring point), the Solimotion can be manually adapted to the application using an
additional manual calibration.
Move the encoding switch to position 2
→ Display of the function number
After 2 seconds: Display of the current signal strength, example:
Simultaneously press the and keys on the device with no movement of bulk solids or
minimum solid flow
→ Automatic calibration is carried out
→ Display of the signal strength with no movement of bulk solids or minimum solid flow
Move the encoding switch to the initial position 0
→ Display of the current signal strength
Optional: Additional manual calibration with no movement of bulk solids or minimum
solid flow (function 4)
Using the manual calibration, which is ideally carried out after the automatic calibration, the
Solimotion can be adapted to the application manually if necessary.
Move the encoding switch to position 4
→ Display of the function number
After 2 seconds: Display of the current signal strength, example:
Press the or key on the device to achieve an increase or reduction of the signal
strength with no movement of bulk solids or minimum solid flow
→ Display of the signal strength with no movement of bulk solids or minimum solid flow
(all 10 LEDs are not lit)
Move the encoding switch to the initial position 0
→ Display of the current signal strength
→
29