Example of bulk solids detection on a conveyor belt
The bulk solids for which movement is to be detected are transported via a conveyor belt. Due
to process fluctuations, the belt is loaded unevenly.
20 Bulk solids detection on a conveyor belt
Default setting of the Solimotion FTR20 (→
47)
Gain (function B)
Detection range (function C)
Signal strength display in this particular example with the default settings:
maximum loading (current output: 15.2 mA)
minimum loading (current output: 8.8 mA)
no loading (current output: 5.6 mA)
A calibration of the switch output and the current output is explained on the following pages,
based on the bulk solids detection on a conveyor belt presented above. These settings are only
examples of a configuration using functions B and C.