Commissioning
Hauser
35
• If bulk solids with bad reflection properties (small signal strength) are detected, the
detection range should be shifted in the direction of smaller signal strengths (high gain).
• If bulk solids with good reflection properties (large signal strength) are detected, the
detection range should be shifted in the direction of larger signal strengths (low gain).
• Adjust the detection range and gain until the display of the signal strength has reliably
exceeded/undershot the upper and lower switch point (→ 31) (switch output)
or the desired output signal has been output (output current).
Overview of the dependency between detection range and gain
Detection range
Maximum gain
• For the calibration it is advisable first to carry out the automatic calibration in function 1 or
2, in order to start with meaningful values.
• The adjustable gain depends on the configured detection range and vice versa.
• Carrying out another automatic calibration (function 1 or function 2) overrides all previous
calibration values.
• When changing devices, the configured values can be directly applied here; another
calibration of the application is not required.