9 Trouble-shooting
Proline Prosonic Flow 90
86
Hauser
9.5
Response of outputs to errors
!
Note!
The failsafe mode of totalizers, current, pulse and frequency outputs can be customized by means
of various functions in the function matrix. You will find detailed information on these procedures
in the “Description of Device Functions” manual.
Positive zero return and failsafe mode:
You can use positive zero return to set the signals of the current, pulse and frequency outputs to
their fallback value, for example when measuring has to be interrupted while a pipe is being
cleaned. This function takes priority over all other device functions; Simulations, for example, are
suppressed.
Failsafe mode of outputs and totalizers
Process/system error is present
Positive zero return is activated
"
Caution!
System or process errors defined as “Notice messages” have no effect whatsoever on the outputs. See the information on
Page 47 ff.
Current output
MINIMUM CURRENT
The current output will be set to the lower value of
the signal on alarm level depending on the setting
selected in the CURRENT SPAN (see the “Descrip-
tion of Device Functions” manual).
MAXIMUM CURRENT
The current output will be set to the higher value of
the signal on alarm level depending on the setting
selected in the CURRENT SPAN (see the “Descrip-
tion of Device Functions” manual).
HOLD VALUE
Measured value display on the basis of the last saved
value preceding occurrence of the fault.
ACTUAL VALUE
Measured value display on the basis of the current
flow measurement. The fault is ignored.
Output signal corresponds to “zero
flow”
Pulse output
FALLBACK VALUE
Signal output
→
no pulses
HOLD VALUE
Last valid value (preceding occurrence of the fault)
is output.
ACTUAL VALUE
Fault is ignored, i.e. standard measured value output
on the basis of ongoing flow measurement.
Output signal corresponds to “zero
flow”
Frequency output
FALLBACK VALUE
Signal output
→
0 Hz
FAILSAFE LEVEL
Output of the frequency specified in the FAILSAFE
VALUE function.
HOLD VALUE
Last valid value (preceding occurrence of the fault)
is output.
ACTUAL VALUE
Fault is ignored, i.e. standard measured value output
on the basis of ongoing flow measurement.
Output signal corresponds to “zero
flow”
Summary of Contents for Prosonic Flow 90
Page 4: ...Brief operating instructions Proline Prosonic Flow 90 4 Endress Hauser...
Page 42: ...4 Wiring Proline Prosonic Flow 90 42 Endress Hauser...
Page 62: ...5 Operation Proline Prosonic Flow 90 62 Endress Hauser...
Page 74: ...7 Maintenance Proline Prosonic Flow 90 74 Endress Hauser...
Page 78: ...8 Accessories Proline Prosonic Flow 90 78 Endress Hauser...
Page 94: ...9 Trouble shooting Proline Prosonic Flow 90 94 Endress Hauser...
Page 108: ...10 Technical data Proline Prosonic Flow 90 108 Endress Hauser...
Page 112: ...Index Proline Prosonic Flow 90 112 Endress Hauser...
Page 114: ...www endress com worldwide BA068D 06 en 11 04 50099981 FM SGLM6 0...