3 Installation
Proline Prosonic Flow 90
18
Hauser
3.3
Installation instructions
3.3.1
Installing tensioning bands (clamp-on)
For W/P sensors DN 50...200
1.
Push one of the supplied threaded bolts on the tensioning band.
2.
Run the tensioning band around the pipe without twisting it and push the end through the
tensioning band lock (make sure that the screw is pushed up).
3.
By hand, make the tensioning band as tight as possible.
4.
Push the screw down and tighten the tensioning band with a screwdriver so that it cannot slip.
5.
If so desired, shorten the tensioning band to the desired length.
"
Caution!
Risk of injury. When shortening the tensioning band, try to avoid sharp edges.
A0001109
Fig. 11:
Tensioning band installation for DN 50...200
For W/P sensors DN 250...4000
The following steps relate to Fig. 12 on Page 19.
1.
Measure the pipe circumference.
Shorten the tensioning band to the pipe circumf10 cm.
"
Caution!
Risk of injury. When shortening the tensioning band, try to avoid sharp edges.
2.
Loop the tensioning band through one of the centering plates supplied with the threaded
bolt (1).
3.
Insert both ends of the tensioning bands down into the openings in the tensioning band lock
(2). Bend back the ends of the tensioning bands.
4.
Interlock both halves of the lock (3). Make sure that there is sufficient space for the tensioning
band to be tightened with the locking screw.
5.
Tighten the tensioning band using a screwdriver (4).
Summary of Contents for Prosonic Flow 90
Page 4: ...Brief operating instructions Proline Prosonic Flow 90 4 Endress Hauser...
Page 42: ...4 Wiring Proline Prosonic Flow 90 42 Endress Hauser...
Page 62: ...5 Operation Proline Prosonic Flow 90 62 Endress Hauser...
Page 74: ...7 Maintenance Proline Prosonic Flow 90 74 Endress Hauser...
Page 78: ...8 Accessories Proline Prosonic Flow 90 78 Endress Hauser...
Page 94: ...9 Trouble shooting Proline Prosonic Flow 90 94 Endress Hauser...
Page 108: ...10 Technical data Proline Prosonic Flow 90 108 Endress Hauser...
Page 112: ...Index Proline Prosonic Flow 90 112 Endress Hauser...
Page 114: ...www endress com worldwide BA068D 06 en 11 04 50099981 FM SGLM6 0...