3 Installation
Proline Prosonic Flow 90
14
Hauser
3.2
Installation conditions
3.2.1
Installation dimensions
The dimensions and the fitting lengths of the sensors and the transmitter are on Page 104 ff.
3.2.2
Installation location
Correct measuring is possible only if the pipe is full.
Avoid
the following installation locations:
• Do not install at the highest point in the run. Risk of air accumulating.
• Do not install directly upstream from an open pipe outlet in a down pipe.
A0001103
Fig. 5:
Installation location
Down pipes
Notwithstanding the above, the installation proposal below permits installation in an open down
pipe. Pipe restrictions or the use of an orifice plate with a smaller cross-section than the nominal
diameter prevent the pipe from running empty while measurement is in progress.
A0001104
Fig. 6:
Installation in a down pipe
1 = Supply tank; 2 = Measuring sensors; 3 = Orifice plate, pipe restriction; 4 = Valve, 5 = Filling tank
1
2
3
4
5
Summary of Contents for Prosonic Flow 90
Page 4: ...Brief operating instructions Proline Prosonic Flow 90 4 Endress Hauser...
Page 42: ...4 Wiring Proline Prosonic Flow 90 42 Endress Hauser...
Page 62: ...5 Operation Proline Prosonic Flow 90 62 Endress Hauser...
Page 74: ...7 Maintenance Proline Prosonic Flow 90 74 Endress Hauser...
Page 78: ...8 Accessories Proline Prosonic Flow 90 78 Endress Hauser...
Page 94: ...9 Trouble shooting Proline Prosonic Flow 90 94 Endress Hauser...
Page 108: ...10 Technical data Proline Prosonic Flow 90 108 Endress Hauser...
Page 112: ...Index Proline Prosonic Flow 90 112 Endress Hauser...
Page 114: ...www endress com worldwide BA068D 06 en 11 04 50099981 FM SGLM6 0...