9 Trouble-shooting
Proline Prosonic Flow 90
80
Hauser
9.2
System error messages
Serious system errors are
always
recognized by the instrument as “Fault message”, and are shown
as a lightning flash (
$
) on the display. Fault messages immediately affect the outputs. Simulations
and measured-value suppression, on the other hand, are classed and displayed as information
messages.
"
Caution!
In the event of a serious fault, a flowmeter might have to be returned to the manufacturer for repair.
The procedures on Page 8 must be carried out before you return a flowmeter to Hauser.
Always enclose a fully completed “Declaration of contamination” form. You will find a preprinted
form at the back of this manual.
!
Note!
The error types listed in the following correspond to the factory settings.
Also observe the information on Page 47 ff. and 86.
Type
Error message / No.
Cause
Remedy / spare part
S = System error
$
= Fault message (
with
an effect on the outputs)
! = Notice message (
without
an effect on the outputs)
No. # 0xx
→
Hardware error
S
$
CRITICAL FAIL.
# 001
Serious device error
Replace the amplifier board. Spare parts
S
$
AMP HW-EEPROM
# 011
Amplifier:
Defective EEPROM
Replace the amplifier board. Spare parts
S
$
AMP SW-EEPROM
# 012
Amplifier:
Error accessing EEPROM data
The EEPROM data blocks in which an
error has occurred are displayed in the
“TROUBLESHOOTING” function.
Press Enter to acknowledge the errors
in question; default values are automat-
ically inserted instead of the errored
parameter values.
!
Note!
The measuring device has to be
restarted if an error has occurred in a
totalizer block (see error No. 111 /
CHECKSUM TOTAL.).
S
$
A / C COMPATIB.
# 051
The I/O board and the amplifier board
are not compatible.
Use only compatible modules and
boards. Check the compatibility of the
modules used.
Check the:
– Spare part set number
– Hardware revision code
S
$
SENSOR DOWN
# 081
Connection between sensor and
transmitter interrupted
– Check the cable connection
between the sensor and the trans-
mitter.
– Check that the sensor connector is
fully screwed in.
– The sensor may be defective.
– Incorrect sensor connected
– A wrong sensor was selected in the
function SENSOR TYPE .
Summary of Contents for Prosonic Flow 90
Page 4: ...Brief operating instructions Proline Prosonic Flow 90 4 Endress Hauser...
Page 42: ...4 Wiring Proline Prosonic Flow 90 42 Endress Hauser...
Page 62: ...5 Operation Proline Prosonic Flow 90 62 Endress Hauser...
Page 74: ...7 Maintenance Proline Prosonic Flow 90 74 Endress Hauser...
Page 78: ...8 Accessories Proline Prosonic Flow 90 78 Endress Hauser...
Page 94: ...9 Trouble shooting Proline Prosonic Flow 90 94 Endress Hauser...
Page 108: ...10 Technical data Proline Prosonic Flow 90 108 Endress Hauser...
Page 112: ...Index Proline Prosonic Flow 90 112 Endress Hauser...
Page 114: ...www endress com worldwide BA068D 06 en 11 04 50099981 FM SGLM6 0...