6 Commissioning
Proline Prosonic Flow 90
68
Hauser
Procedure (determining data for sensor installation)
You can use the following tables to select and configure, in the correct order, the functions required
to install the sensor:
• “clamp-on” sensor installation
• “Insertion” sensor installation
!
Note!
You must enter a valid release code before device parameters can be changed or activated.
The code (factory setting = 90) is entered by means of the corresponding matrix cell.
“Clamp-on” sensor installation
Procedure
Selection - Input - display
Local display (Quick Setup)
▼
ToF Tool - Fieldtool Package
▼
▼
→
BASIC FUNCTION
→
SENSOR DATA
→
SENSOR PARAM.
→
BASIC FUNCTION
→
SENSOR DATA
→
SENSOR PARAM.
Type of measurement
MEASUREMENT
MEASUREMENT
Sensor type
SENSOR TYPE
SENSOR TYPE
Sensor configuration
SENSOR CONFIGURATION
SENSOR CONFIGURATION
▼
→
BASIC FUNCTION
→
PROCESS PARAM.
→
PIPE DATA
→
BASIC FUNCTION
→
PROCESS PARAM.
→
PIPE DATA
Standard pipe selection
PIPE STANDARD
PIPE STANDARD
Nominal diameter pipe
NOMINAL DIAMETER
NOMINAL DIAMETER
Pipe material
PIPE MATERIAL
PIPE MATERIAL
Pipe sound velocity
SOUND VELOCITY PIPE
SOUND VELOCITY PIPE
Pipe circumference
CIRCUMFERENCE
CIRCUMFERENCE
Pipe diameter
PIPE DIAMETER
PIPE DIAMETER
Wall thickness
WALL THICKNESS
WALL THICKNESS
Liner material
LINER MATERIAL
LINER MATERIAL
Liner sound velocity
SOUND VELOCITY LINER
SOUND VELOCITY LINER
Liner thickness
LINER THICKNESS
LINER THICKNESS
▼
→
BASIC FUNCTION
→
PROCESS PARAM.
→
LIQUID DATA
→
BASIC FUNCTION
→
PROCESS PARAM.
→
LIQUID DATA
Liquid in the pipe
LIQUID
LIQUID
Liquid temperature
TEMPERATURE
TEMPERATURE
Liquid sound velocity
SOUND VELOCITY LIQUID
SOUND VELOCITY LIQUID
▼
→
BASIC FUNCTION
→
SENSOR DATA
→
SENSOR PARAM.
→
BASIC FUNCTION
→
SENSOR DATA
→
SENSOR PARAM.
Didsplay sensorposition
(for sensor installation)
POSITION SENSOR
POSITION SENSOR
Summary of Contents for Prosonic Flow 90
Page 4: ...Brief operating instructions Proline Prosonic Flow 90 4 Endress Hauser...
Page 42: ...4 Wiring Proline Prosonic Flow 90 42 Endress Hauser...
Page 62: ...5 Operation Proline Prosonic Flow 90 62 Endress Hauser...
Page 74: ...7 Maintenance Proline Prosonic Flow 90 74 Endress Hauser...
Page 78: ...8 Accessories Proline Prosonic Flow 90 78 Endress Hauser...
Page 94: ...9 Trouble shooting Proline Prosonic Flow 90 94 Endress Hauser...
Page 108: ...10 Technical data Proline Prosonic Flow 90 108 Endress Hauser...
Page 112: ...Index Proline Prosonic Flow 90 112 Endress Hauser...
Page 114: ...www endress com worldwide BA068D 06 en 11 04 50099981 FM SGLM6 0...