3 Installation
Proline Prosonic Flow 90
30
Hauser
5.
Mark the sensor distance (a) from the middle line starting from the first drillhole.
6.
Project the middle line to the back of the pipe and draw it on.
A0001126
Fig. 26:
Installing the measuring sensors, steps 5 and 6
7.
Mark the drillhole on the middle line on the back of the pipe.
8.
Cut out the second drillhole and prepare the holes for welding of the sensor holder
(deburr, clean, etc.).
A0001127
Fig. 27:
Installing the measuring sensors, steps 7 and 8
9.
Insert the sensor holders into the two drill holes. To adjust the weld-in depth, both sensor
holders can be fixed with the special tool for insertion depth regulation (optional) and then
aligned using the tie rod. The sensor holder must be flush with the inner side of the pipe.
Now pinpoint both sensor holders.
!
Note!
To align the tie rod, two bearing shells must be screwed onto the sensor holders.
A000128
Fig. 28:
Installing the measuring sensors, step 9
a
Summary of Contents for Prosonic Flow 90
Page 4: ...Brief operating instructions Proline Prosonic Flow 90 4 Endress Hauser...
Page 42: ...4 Wiring Proline Prosonic Flow 90 42 Endress Hauser...
Page 62: ...5 Operation Proline Prosonic Flow 90 62 Endress Hauser...
Page 74: ...7 Maintenance Proline Prosonic Flow 90 74 Endress Hauser...
Page 78: ...8 Accessories Proline Prosonic Flow 90 78 Endress Hauser...
Page 94: ...9 Trouble shooting Proline Prosonic Flow 90 94 Endress Hauser...
Page 108: ...10 Technical data Proline Prosonic Flow 90 108 Endress Hauser...
Page 112: ...Index Proline Prosonic Flow 90 112 Endress Hauser...
Page 114: ...www endress com worldwide BA068D 06 en 11 04 50099981 FM SGLM6 0...