Proline Prosonic Flow 90
5 Operation
Hauser
57
5.5.5
Device status / Error messages
You can read the extended device status, in this case, current error messages, via Command “48”.
The command delivers information which are partly coded in bits (see table below).
!
Note!
You can find a detailed explanation of the device status and error messages and their elimination on
Page 80 ff.!
Byte
Bit
Error no.
Short error description (
0
0
001
Serious device error
1
011
Measuring amplifier has faulty EEPROM
2
012
Error when accessing data of the measuring amplifier EEPROM
3
not assigned
–
4
not assigned
–
5
not assigned
–
6
not assigned
–
7
not assigned
–
1
0
not assigned
–
1
not assigned
–
2
not assigned
–
3
not assigned
–
4
not assigned
–
5
not assigned
–
6
not assigned
–
7
not assigned
–
2
0
not assigned
–
1
081
Connection (downwards) sensor/transmitter interrupted
2
not assigned
–
3
not assigned
–
4
084
Connection (upwards) sensor/transmitter interrupted
5
not assigned
–
6
not assigned
–
7
not assigned
–
3
0
not assigned
–
1
not assigned
–
2
not assigned
–
3
111
Totalizer checksum error
4
121
The I/O- and the amplifier board are not compatible.
5
not assigned
–
6
not assigned
–
7
not assigned
–
Summary of Contents for Prosonic Flow 90
Page 4: ...Brief operating instructions Proline Prosonic Flow 90 4 Endress Hauser...
Page 42: ...4 Wiring Proline Prosonic Flow 90 42 Endress Hauser...
Page 62: ...5 Operation Proline Prosonic Flow 90 62 Endress Hauser...
Page 74: ...7 Maintenance Proline Prosonic Flow 90 74 Endress Hauser...
Page 78: ...8 Accessories Proline Prosonic Flow 90 78 Endress Hauser...
Page 94: ...9 Trouble shooting Proline Prosonic Flow 90 94 Endress Hauser...
Page 108: ...10 Technical data Proline Prosonic Flow 90 108 Endress Hauser...
Page 112: ...Index Proline Prosonic Flow 90 112 Endress Hauser...
Page 114: ...www endress com worldwide BA068D 06 en 11 04 50099981 FM SGLM6 0...