Proline Prosonic Flow 90
9 Trouble-shooting
Hauser
81
S
$
SENSOR UP
# 084
Connection between sensor and
transmitter interrupted
– Check the cable connection
between the sensor and the
transmitter.
– Check that the sensor connector is
fully screwed in.
– The sensor may be defective.
– Incorrect sensor connected
– A wrong sensor was selected in the
function SENSOR TYPE .
No. # 1xx
→
Software error
S
$
CHECKSUM TOT.
# 111
Totalizer checksum error
1.
Restart the measuring device
2.
Replace the amplifier board if
necessary.
Spare parts
S
$
A / C COMPATIB.
# 121
Due to different software versions,
I/O board and amplifier board are
only partially compatible (possibly
restricted functionality).
!
Note!
– The indication on the display as
notice message appears only for 30
seconds (with listing in “Previous
system condition” function).
– This condition can occur if only one
electronics board has been
exchanged; the extended software
functionality is not available. The
previously existing software func-
tionality is still working and the
measurement possible.
Module with lower software version
has either to be actualizied by ToF Tool
- Fieldtool Package with the required
software version or the module has to
be replaced.
Spare parts
No. # 3xx
→
System range limits exceeded
S
$
STACK CUR.OUT n
# 339...342
The temporarily buffered flow por-
tions (measuring mode for pulsating
flow) could not be cleared or output
within 60 seconds.
1.
Change the upper or lower limit
setting, as applicable.
2.
Increase or reduce flow, as applica-
ble.
Recommendations in the event of fault
category = FAULT MESSAGE (
$
):
– Configure the fault response of the
output to “ACTUAL VALUE”
(see Page 86), so that the temporary
buffer can be cleared.
– Clear the temporary buffer by the
measures described under Item 1.
S
$
STACK FRQ.OUT n
# 343...346
S
$
STACK PULSE n
# 347...350
The temporarily buffered flow por-
tions (measuring mode for pulsating
flow) could not be cleared or output
within 60 seconds.
1.
Increase the setting for pulse
weighting
2.
Increase the max. pulse frequency,
if the totalizer can handle a higher
number of pulses.
3.
Increase or reduce flow.
Recommendations in the event of fault
category = FAULT MESSAGE (
$
):
– Configure the fault response of the
output to “ACTUAL VALUE”
(see Page 86), so that the temporary
buffer can be cleared.
– Clear the temporary buffer by the
measures described under Item 1.
Type
Error message / No.
Cause
Remedy / spare part
Summary of Contents for Prosonic Flow 90
Page 4: ...Brief operating instructions Proline Prosonic Flow 90 4 Endress Hauser...
Page 42: ...4 Wiring Proline Prosonic Flow 90 42 Endress Hauser...
Page 62: ...5 Operation Proline Prosonic Flow 90 62 Endress Hauser...
Page 74: ...7 Maintenance Proline Prosonic Flow 90 74 Endress Hauser...
Page 78: ...8 Accessories Proline Prosonic Flow 90 78 Endress Hauser...
Page 94: ...9 Trouble shooting Proline Prosonic Flow 90 94 Endress Hauser...
Page 108: ...10 Technical data Proline Prosonic Flow 90 108 Endress Hauser...
Page 112: ...Index Proline Prosonic Flow 90 112 Endress Hauser...
Page 114: ...www endress com worldwide BA068D 06 en 11 04 50099981 FM SGLM6 0...