Proline Prosonic Flow 90
5 Operation
Hauser
55
35
Write measuring range of primary
process variable
Access = write
Write the desired measuring range:
– Byte 0: HART unit code for the primary process
variable
– Byte 1
–
4: Upper range, value for 20 mA
– Byte 5
–
8: Lower range, value for 4mA
Factory setting:
Primary process variable = Volume flow
!
Note!
• You can set or change the assignment of device
variables to process variables using Command 51.
• If the HART unit code is not the correct one for the
process variable, the device will continue with the
last valid unit.
The currently set measuring range is shown as the
response:
– Byte 0: HART unit code for the set measuring range
of the primary process variable
– Byte 1
–
4: Upper range, value for 20 mA
– Byte 5
–
8: Lower range, value for 4mA
!
Note!
Manufacturer-specific units are represented using the
HART unit code “240”.
38
Device status reset “Configuration
changed”
Access = write
none
none
40
Simulate output current of
primary process variable
Access = write
Simulation of the desired output current of the primary
process variable. An entry value of 0 exits the simula-
tion mode:
Byte 0
–
3: Output current in mA
Factory setting:
Primary process variable = Volume flow
!
Note!
You can set the assignment of device variables to
process variables using Command 51.
The momentary output current of the primary process
variable is displayed as a response:
Byte 0
–
3: Output current in mA
42
Perform device reset
Access = write
none
none
44
Write unit of primary process
variable
Access = write
Set unit of primary process variable Only unit which
are suitable for the process variable are transferred to
the device:
Byte 0: HART unit code
Factory setting:
Primary process variable = Volume flow
!
Note!
• If the written HART unit code is not the correct one
for the process variable, the device will continue
with the last valid unit.
• If you change the unit of the primary process
variable, this has no impact on the system units.
The current unit code of the primary process variable is
displayed as a response:
Byte 0: HART unit code
!
Note!
Manufacturer-specific units are represented using the
HART unit code “240”.
48
Read additional device status
Access = read
none
The device status is displayed in extended form as the
response:
Coding: see table on Page 57
Command No.
HART command / Access type
Command data
(numeric data in decimal form)
Response data
(numeric data in decimal form)
Summary of Contents for Prosonic Flow 90
Page 4: ...Brief operating instructions Proline Prosonic Flow 90 4 Endress Hauser...
Page 42: ...4 Wiring Proline Prosonic Flow 90 42 Endress Hauser...
Page 62: ...5 Operation Proline Prosonic Flow 90 62 Endress Hauser...
Page 74: ...7 Maintenance Proline Prosonic Flow 90 74 Endress Hauser...
Page 78: ...8 Accessories Proline Prosonic Flow 90 78 Endress Hauser...
Page 94: ...9 Trouble shooting Proline Prosonic Flow 90 94 Endress Hauser...
Page 108: ...10 Technical data Proline Prosonic Flow 90 108 Endress Hauser...
Page 112: ...Index Proline Prosonic Flow 90 112 Endress Hauser...
Page 114: ...www endress com worldwide BA068D 06 en 11 04 50099981 FM SGLM6 0...