9 Trouble-shooting
Proline Prosonic Flow 90
82
Hauser
S
!
RANGE CUR.OUT n
# 351...354
Current output:
Flow is out of range.
– Change the upper or lower range
values, as applicable.
– Increase or reduce flow, as applica-
ble.
S
!
RANGE FRQ.OUT n
# 355...358
Frequency output:
Flow is out of range.
– Change the upper or lower range
values, as applicable.
– Increase or reduce flow, as applica-
ble.
S
!
RANGE PULSE n
# 359...362
Pulse output:
Pulse output frequency is out of range.
1.
Increase the setting for pulse
weighting
2.
When selecting the pulse width,
choose a value that can still be
processed by a connected counter
(e.g. mechanical counter, PLC
etc.).
Determine the pulse width:
– Variant 1: Enter the minimum
duration that a pulse must be
present at the connected coun-
ter to ensure its registration.
– Variant 2: Enter the maximum
(pulse) frequency as the half
“reciprocal value” that a pulse
must be present at the con-
nected counter to ensure its
registration.
Example:
The maximum input frequency of
the connected counter is 10 Hz.
The pulse width to be entered is:
3.
Reduce flow.
S
$
SIGNAL TOO LOW
# 391
Attenuation of acoustic measurement
section too high.
– Check to see if the coupling fluid
must be renewed.
– It is possible that the fluid indicates
too much attenuation.
– It is possible that the pipe indicates
too much attenuation.
– Check the sensor spacing
(Installation dimensions).
– Reduce the number of traverses if
possible.
No. # 5xx
→
Application error
S
!
SW.-UPDATE ACT.
# 501
New amplifier or communication (I/O
module) software version is loaded.
Currently no other functions are pos-
sible.
Wait until the procedure is finished.
The device will restart automatically.
S
!
UP-/DOWNLOAD ACT
# 502
Up- or downloading the device data
via configuration program. Currently
no other functions are possible.
Wait until the procedure is finished.
S
$
INITIAL. RUN
# 591
Initialisation is running.
All outputs are set to 0.
Wait until the procedure is complete.
Type
Error message / No.
Cause
Remedy / spare part
1
2 10 Hz
⋅
----------------------
50 ms
=
Summary of Contents for Prosonic Flow 90
Page 4: ...Brief operating instructions Proline Prosonic Flow 90 4 Endress Hauser...
Page 42: ...4 Wiring Proline Prosonic Flow 90 42 Endress Hauser...
Page 62: ...5 Operation Proline Prosonic Flow 90 62 Endress Hauser...
Page 74: ...7 Maintenance Proline Prosonic Flow 90 74 Endress Hauser...
Page 78: ...8 Accessories Proline Prosonic Flow 90 78 Endress Hauser...
Page 94: ...9 Trouble shooting Proline Prosonic Flow 90 94 Endress Hauser...
Page 108: ...10 Technical data Proline Prosonic Flow 90 108 Endress Hauser...
Page 112: ...Index Proline Prosonic Flow 90 112 Endress Hauser...
Page 114: ...www endress com worldwide BA068D 06 en 11 04 50099981 FM SGLM6 0...