8 Troubleshooting
Mycom S CLM 153
88
EHauser
8.1.1
Error number list: Troubleshooting and configuration
In the following error list, you can find a description of all the error numbers occurring.
For each error number there is also data on whether the factory setting (= Fact.)
of this error triggers
•
an alarm,
•
an error current or
•
cleaning.
To enter the error list, proceed as follows:
!
Note!
•
Please process the errors as shown in Field H5 (Alarm menu) on page 45.
•
The second column shows the assignment acc. to NAMUR work sheet NA64 (failure,
maintenance, function control).
⇒
⇒
Diag
Edit
Next
Select
E
Hold
Error list
Error log
Operating log
Calibration log
Service
3.52 mS/cm
Error no.
NAMUR
class
Error message
Possible causes / measures
Alarm
contact
Error
current
Automatic
cleaning
start
Fact.
User
Fact.
User
Fact.
User
E001
Failure
Memory error
Switch instrument off and back on.
Corrective maintenance in the factory,
if necessary.
yes
no
–
–
E002
Failure
User data invalid
yes
no
–
–
E003
Failure
No valid configuration
E004
Failure
Erro invalid hardware or software
The new software cannot recognise
the module.
E006
Failure
Transmitter 2 error
Test with new transmitter.
yes
no
–
–
E007
Failure
Transmitter 1 error
yes
no
–
–
E008
Failure
Sensor error 1
Check sensor and sensor connection
(chap. 7.1.5 / chap. 7.1.6 or by E+H
service)
yes
no
no
E009
Failure
Sensor error 2
yes
no
no
E010
Failure
Temperature sensor 1 defective
Check temperature sensor and
connections; if necessary, check
transmitter with temperature
simulator.
yes
no
no
E011
Failure
Temperature sensor 2 defective
yes
no
no
E019
Failure
Delta exceeded failure
Check single measured values
(input 1 / input 2) for plausibility
yes
no
–
–
E025
Failure
Limit Airset Offset CH 1 exceeded Carry out airset again (only inductive;
at air) or replace sensor. Clean and
dry sensor before airset.
E026
Failure
Limit Airset Offset CH 2 exceeded
E034
Failure
Cell constant sensor 1 exceeded
Clean sensor and recalibrate; if
necessary, check sensor, line and
connections.
yes
no
–
–
E035
Failure
Cell constant sensor 1 undershot
yes
no
–
E036
Failure
Cell constant sensor 2 exceeded
yes
no
–
E037
Failure
Cell constant sensor 2 undershot
yes
no
–