10 Technical data
Mycom S CLM 153
102
EHauser
Resistance input
(active, optional,
only with non-Ex)
Digital inputs
1
: acc. to IEC 746-1, under nominal operating conditions
10.2
Output
Output signal
Conductivity, resistance, concentration, difference, efficiency, temperature
Current outputs
1
: acc. to IEC 746-1, under nominal operating conditions
Auxiliary voltage output
(for digital inputs
E1-E3)
Limit value and
alarm functions
Resistance ranges (software switchable)
0 ... 1 k
Ω
0 ... 10 k
Ω
Measured error
1
max. 1% of measuring range
Input voltage
10 ... 50 V
Internal resistance
R
i
= 5 k
Ω
Current range
0 / 4 ... 20 mA
Error current
2.4 mA or 22 mA
Measured error
1
< 0.2 % of
current range maximum
Transmission characteristic
Linear, bilinear
Output distribution, temperature
17 ... 170 °C
Output distribution, minimum (maximum up to 100% of the measuring range):
For conductivity measurement for measured
values in the range of:
0 ... 19.99 µS/cm:
20 ... 199.9 µS/cm:
200 ... 1999 µS/cm:
2 ... 19.99 mS/cm:
20 ... 2000 mS/cm:
2 µS/cm
20 µS/cm
200 µS/cm
2 mS/cm
20 mS/cm
For resistance measurement for measured
values in the range of:
0 ... 199.9 k
Ω
·cm:
200 ... 1999 k
Ω
·cm:
2 ... 19.99 M
Ω
·cm:
20 ... 200 M
Ω
·cm:
20 k
Ω
·cm
200 k
Ω
·cm
2.0 M
Ω
·cm
20 M
Ω
·cm
For concentration measurement
No minimum spacing
Active current output (only non-Ex): Load
Max. 600
Ω
Passive current output:
Input voltage range
6 ... 30 V
Voltage
15 V DC
Output current
max. 9 mA
Setpoint adjustments
0 ... 100% of display range
Hysteresis for switching contacts
1 ... 10% of display range
Alarm delay entry
0 ... 6000 s