Mycom S CLM 153
6 Commissioning
Hauser
79
6.4.20
Calibration
The calibration can be protected with the maintenance and the specialist codes. No
calibration can be carried out at the display level (compare with page 33).
Procedure:
1.
Move assembly to service position (when a rectractable assembly is used).
2.
Remove sensor.
3.
Clean sensor before calibration.
Calibration can be performed in two different ways:
•
By measuring with a known conductivity (Fields xx to yy) in a calibration solution
•
By entering the precise cell constant of the conductivity sensor (Field A5, p. 31).
!
Note!
•
If automatic temperature compensation is selected for calibration (s. page 27), the
corresponding temperature sensor must also be immersed in the calibration solution.
•
The instrument switches automatically to Hold (factory setting) whenever it is
calibrated.
•
Cancel calibration by pressing the "MEAS" key.
•
If you confirm this with "yes, cancel cal.", you return to the
measurement mode.
•
If you select "no", calibration is continued.
3.52 mS/cm
Cal
Edit
E
Next
Cancel Calib.
yes, Cancel Calib.
yes,
no