Mycom S CLM 153
6 Commissioning
Hauser
27
T11
ATC CH1
ATC CH2
MTC
MTC+Temp
Select temperature compensation channel
1/2
ATC:
Automatic temperature compensation
via temperature sensor
MTC:
Temperature compensation via manual
value entry
MTC+Temp:
Temperature compensation via
manual temperature entry. The displayed
temperature, however, is measured by the
temperature sensor.
T12
Pt 100
Pt 1000
NTC 30k
Select temperature sensor channel1
T13
Temp.comp
Alpha
value:
Actual
temp.:
Offset:
linear
2.1%/K
25.0°C
0.0°C
Temperature compensation input 1
Temp. value
: Automatic temperature
compensation with a temp. sensor input 1 or
input 2
Temp. comp
.: Selection of the temperature
compensation – none, linear, NaCl, Table 1 to
4
Alpha value:
Entry of the conductivity
coefficient
α
.
Actual temp
.: The measured temperature is
displayed.
Offset
: Temperature difference between
measured and output temperature
(–10 ... +10 °C).
T14
ATC CH1
ATC CH2
MTC
MTC+Temp
Select temperature compensation
channel 2
(only two circuit)
T15
Pt 100
Pt 1000
NTC 30k
Select temperature sensor input 2
(only two circuit)
T16
Temp.value
Temp.comp
.:
Alpha
value:
Actual
temp.:
Offset:
Input 1
none
2.1%/K
25.0°C
0.0°C
Temperature compensation input 2
(only two circuit)
Temp. value
: Automatic temperature
compensation with a temp. sensor input 1 or
input 2
Temp. comp
.: Selection of the temperature
compensation – none, linear, NaCl, Table 1 to
4
Alpha value:
Entry of the conductivity
coefficient
α
.
Actual temp
.: The measured temperature is
displayed.
Offset
: Temperature difference between
measured and output temperature
(–10 ... +10 °C).
CODE
DISPLAY
CHOICE
(default = bold)
INFO
E
Hold
Param
Edit
3
mS/cm
.52
Temp.comp.CH1
Next
ATC CH1
ATC CH2
MTC
MTC+Temp
E
Hold
Param
Edit
3
mS/cm
.52
Pt1000
Pt100
NTC30
Temp.sensor K1
Next
E
Hold
Param
Edit
3
mS/cm
.52
02.10%/K
0.0°C
025.0°C
linear
Compensation CH1
Next
Temp.comp.
Alpha value
Offset
Actual temp.
E
Hold
Param
Edit
3
mS/cm
.52
Temp.comp.CH2
Next
ATC CH1
ATC CH2
MTC
MTC+Temp
E
Hold
Param
Edit
3
mS/cm
.52
Pt1000
Pt100
NTC30
Temp.sensor K2
Next
E
Hold
Param
Edit
3
mS/cm
.52
K1
02.10%/K
0.0°C
025.0°C
linear
Compensation K2
Next
Temp.comp.
Temp.value
Alpha value
Offset
Actual temp.