5 Operation
Mycom S CLM 153
20
EHauser
5.1.4
Data logger
In the CLM 153, you have two data loggers available. With these data loggers you can
•
record a parameter with 500 sequential measuring points or
•
two parameters each with 250 sequential measuring points.
To be able to use the function, activate the data logger(s) in the "PARAM" menu
➞
"Set
up 2"
➞
"Data Logger" (s. page 52). The function is active immediately.
You can view the measured values by scrolling through the different measuring menus
(see above).
– The current measured values are recorded in Record mode.
– In Scroll mode, you can open saved data by stating the date and time. You can find
this under the item "PARAM"
➞
"Set up 2"
➞
"Data logger".
5.1.5
Operation access authorisation
To protect the transmitter against an unintended or undesired change in the
configuration and calibration data, functions can be protected using four-digit access
codes.
Access authorisation has the following levels:
Display level (accessible without a code):
The complete menu can be viewed. The configuration cannot be altered. No calibration
is possible. On this level, only the control parameters for new processes can be
changed in the "DIAG" menu branch.
Maintenance code
Maintenance level (can be protected by the maintenance code):
This code allows access to the calibration menu.
Use this code to operate the temperature compensation item. The factory functions and
the internal data can be viewed.
Factory setting Code = 0000, i.e. the levels are not protected.
In case you have mislaid/forgotten the supplied maintenance code, contact your E+H
service office.
Specialist code
Specialist level (can be protected by the specialist code):
All menus can be accessed and changed.
Factory setting Code = 0000, i.e. the levels are not protected.
In case you have mislaid/forgotten the supplied specialist code, contact your E+H
service.
To activate the codes (= functions locked) see the item "PARAM"
➞
"Set up 1"
➞
"Access
codes" (s. page 33).
Enter your desired codes here. If the code is activated, you can
only edit the protected areas with the rights mentioned above.
Record mode
Scroll mode
0.00
1000
PS1
mS/cm 1
Measure
Select
0.00
3.52
12 15 35
01 04 2001
1000
PS1
mS/cm 1
Measure
Select