6 Commissioning
Mycom S CLM 153
40
EHauser
Create alpha table:
GB1
Table 1
Table 2
Table 3
Table 4
Select table
Select a table to edit.
GBB2
01
(1 ... 10)
Entry of the number of support points (value pairs)
Value pair: Temperature and conductivity coefficient
α
.
GBB3
°C
000.0
%/K
00.00
Entry of the value pairs
Enter temperature and conductivity coefficient,
(number of value pairs required = number of support
points desired in Field GBB2).
GBB4
OK
Delete element(s)
Selection:
Are the value pairs OK or do you want to delete
elements?
GBB5
°C
020.0 °C
025.0 °C
%/K
02.00
04.00
Delete:
Select the rows to be deleted, delete them with
T
and
confirm this with "E".
GBB6
Valid table
Note in display:
The table is active after confirmation by pressing "E".
Cancel by pressing "PARAM".
Reference temperature:
GBC1
For laboratory
measurement:
25.0 °C
(-35 ... +250 °C)
Entry of the reference temperature
to which the medium temperature should be
compensated. Here, enter the temperature at which the
α
value was determined, (you can take this from the
data sheets from which you also took the
α
value).
CODE
CHOICE
(default = bold)
INFO