6 Commissioning
Mycom S CLM 153
42
EHauser
Fig. 17:
Measuring data in the event of variable process temperature
Fig. 18:
Measuring data in the event of constant process temperature
!
Note!
•
Danger of inaccuracy.
Please ensure that the concentrations and temperatures
measured for your samples also correspond to the measuring range of the process.
If the measured values of the process are outside the range of your sample values,
this considerably reduces the level of accuracy and the device generates an error
message.
•
You can work without error messages from the start of measuring range if you enter
an additional triple value set with 0 µS/cm and 0% for each temperature used, with
ascending characteristic (see diagrams above).
Enter the three characteristic quantities (triple value set with conductivity, temperature
and concentration) for each measured sample into the field Z5.
Concentration
C
Concentration
C
C
min
C
min
T
1
T
1
T
2
T
2
T
3
T
3
C
max
C
max
Conductivity
Conductivity
Meas. range
Meas. range
Meas. point
Meas. point
C07-CLM153xx-05-06-00-en-004.EPS
Concentration
C
T constant
C
min
C
ma x
Concentration
C
T constant
C
min
C
ma x
Conductivity
Meas. range
Conductivity
Meas. range
C07-CLM153xx-05-06-00-en-005.EPS