6 Commissioning
Mycom S CLM 153
36
EHauser
6.4.5
Set up 1 – Relays
To enter the menu, proceed as follows:
⇒
⇒
PARAM
pH
7.00
Param
Parametrieren
Grundeinstellung
Sonderfunktionen
Handbedienung
Quick
Setup
Edit
Weiter E
Hold
Param
Settings
Set up 1
Set up 2
Manual operation
First start up
Edit
Next E
3.52 mS/cm
Param
Set up 1
Sensor input
Display
Access codes
Current output
Relays
Edit
Next E
Hold
3.52 mS/cm
CODE
CHOICE
(default = bold)
INFO
F1
NAMUR:
Relay 1:
Relay 2:
Relay 3:
Relay 4:
Relay 5:
off
free
free
free
free
free
Contact functions
Depending on the equipment available, you can assign
the function of up to 5 relays here. When you switch on
the status message acc. to NAMUR NA64, relays 1 and 2
are assigned and are not available for another function,
(compare page 14).
Selection: Free / Controller / LC1 / LC2 / LC3 / LC4 / LC5
/ CCW / CCC
Controller:
Controller control using relay
LC
: Limit contactor function
CCW
: ChemoClean water. Water supply for the
ChemoClean function.
CCC
: ChemoClean Cleaner. Cleaner supply for the
ChemoClean function.(Together, CCC and CCW form the
"ChemoClean" function. You can find information on
ChemoClean on page 70).
The limit value contacts are configured in the menu
"PARAM"
➞
"Set up 2"
➞
"Limit switch".
The controller contacts are configured in the menu
"PARAM"
➞
"Set up 2"
➞
"Controller settings".
!
Note!
•
Danger of data loss.
If the controller is already
completely configured for output via relay and you
reduce the number of relays assigned to the controller,
the
entire
controller configuration (s. page 54) is reset
to the default values.
• If you change the relay assignment for the controllers,
you must use the controller menu (s. page 54) to
reassign a relay all the functions selected there.
Example
: Relays 4 and 5 are assigned to the
controller and you change the controller assignment to
relays 5 and 6 (number of relays remains 2)
(no data loss, providing the number of assigned relays
is not reduced!).
• You can only activate NAMUR if the required relays 1
and 2 (compare page 14) are free.
• If you want to use the USP function, first select the limit
contactor function n the menu "PARAM"
➞
"Set up 1"
➞
"Contacts" for a relay and configure this in the menu
"PARAM"
➞
"Set up 2"
➞
"Limit switch" for USP.
Activate the contact with "Function: on".