![Endress+Hauser Mycom S CLM 153 Operating Instructions Manual Download Page 89](http://html1.mh-extra.com/html/endress-hauser/mycom-s-clm-153/mycom-s-clm-153_operating-instructions-manual_2408551089.webp)
Mycom S CLM 153
8 Troubleshooting
Hauser
89
E038
Mainte-
nance
Delta limit exceeded
Check single measured values (input
1 / input 2) for plausibility Measuring
can still continue until failure error
E019 occurs.
yes
no
–
–
E046
Failure
Installation factor CH 1exceeded
Check pipe diameter, clean sensor
and carry out calibration again (only
for inductive).
E047
Failure
Installation factor CH1 undershot
E048
Mainte-
nance
Installation factor CH 2 exceeded
yes
no
–
–
E049
Mainte-
nance
Installation factor CH 2 undershot
yes
no
–
–
E053
Failure
Actuator error
E054
Mainte-
nance
Dosing time monitor
Feedback to controller defective
yes
no
–
–
E055
Failure
Display / measuring range of
main parameter 1 undershot
Immerse sensor in conductive
medium.
For inductive: Carry out airset.
As long as the sensor is out of the
medium during Airset, an error code
referring to measuring range violation
can be ignored.
yes
no
no
E056
Failure
Display / measuring range of
main parameter 2 undershot
yes
no
no
E057
Failure
Display / measuring range of
main parameter 1 exceeded
yes
no
no
E058
Failure
Display / measuring range of
main parameter 2 exceeded
yes
no
no
E059
Failure
Temperature measuring range 1
undershot
Temperature sensor defective;
Sensor line interrupted or short-
circuited;
Incorrect sensor type selected
Simulation s. chap. 7.1.3 /
chap. 7.1.4.
yes
no
no
E060
Failure
Temperature measuring range 2
undershot
yes
no
–
–
E061
Failure
Temperature measuring range 1
exceeded
yes
no
no
E062
Failure
Temperature measuring range 2
exceeded
yes
no
–
–
E063
Failure
Current range current output 1
undershot
Measured value outside specified
current range:
Check measured value for plausibility,
if necessary adjust current output
assignment 0/4 mA and/or 20 mA.
yes
no
no
E064
Failure
Current range current output 1
exceeded
yes
no
no
E065
Failure
Current range current output 2
undershot
yes
no
–
–
E066
Failure
Current range current output 2
exceeded
yes
no
–
–
E067
Mainte-
nance
Setpoint controller / LS 1
exceeded
Dosing devices defective;
chemical supply empty;
measured value incorrect
➞
check
for plausibility and function;
Incorrect control direction set;
incorrect contact assigned;
Incorrect control function assigned
yes
no
–
–
E068
Mainte-
nance
Setpoint controller / LS 2
exceeded
yes
no
–
–
E069
Mainte-
nance
Setpoint controller / LS 3
exceeded
yes
no
–
–
E070
Mainte-
nance
Setpoint controller / LS 4
exceeded
yes
no
no
E071
Mainte-
nance
Setpoint controller / LS 5
exceeded
yes
no
no
Error no.
NAMUR
class
Error message
Possible causes / measures
Alarm
contact
Error
current
Automatic
cleaning
start
Fact.
User
Fact.
User
Fact.
User