Mycom S CLM 153
11 Appendix
Hauser
113
CAL
C07-CLM153xx-19-06-08-en-010.EPS
Selection
1
2
Abort Calibration
(only 2-circuit)
Sensor
Sensor
If you press the MEAS
key, a message will
appear asking if you
want to abort the
calibration.
= Code entry required
Airset
Cell constant
Installation factor
Residual coupling
start calibration
current meas.value
0.0µS/cm
Display residual
coupling
42µS/cm
Display calibration
status
o.k.
Calibration status
is displayed
o.k.
Calibration status
is displayed
o.k.
Calibration
Accept
Reject
Recalibrate
Calibration
Accept
Reject
Recalibrate
Press [E]
Press [E]
Press [E]
Calibration
Accept
Reject
Recalibrate
Alpha value of
calibration solution
2.10 %/K
(0.00...20.00%/K)
Alpha value of
calibration solution
2.10 %/K
(0.00...20.00%/K)
Enter calibration
temperature
025.0°C
(-10.0...150.0°C)
Enter calibration
temperature
025.0°C
(-10.0...150.0°C)
Calculated cell
constant is displayed
0,1...9,99 1/cm
Calculated install.
factor is displayed
0,1...9,99 1/cm
Enter conductivity
value of cal.solution
current meas.value
(0.0.... 9999mS/cm)
Enter conductivity
value of cal.solution
current meas.value
(0.0.... 9999mS/cm)