6 Commissioning
Mycom S CLM 153
74
EHauser
6.4.19
Diagnosis
To enter the menu, proceed as follows:
⇒
⇒
Diag
Edit
Next
Select
E
Hold
Error list
Error log
Operating log
Calibration log
Service
3.52 mS/cm
CODE
CHOICE
(default = bold)
INFO
U
Error list
Error log
Operation log
Calibration log
service
Error list
: Displays the current active errors.
(Complete error list with description s. page 88)
Error log
: Lists the last 30 signalled errors with date and
time.
Operation log
(service code necessary)
:
Lists the last
30 registered operating steps with date and time.
Calibration log
: Lists the last 30 calibrations undertaken
with date and time.
!
Note!
• Use the arrow keys to scroll through the lists.
• Leave the lists by pressing "E".
Service:
Y
Factory settings
Simulation
Instrument check
Reset
DAT handling
Instrument version
ChemoClean
Factory function
Selection for service diagnosis
Factory settings
: Different data groups can be reset to
the factory settings.
Simulation
: The transmitter behaviour can be simulated
after entering various parameters.
Instrument check
: The instrument functions
(display, keys, etc.) can be tested individually.
Reset
: Device reset ("warm start")
DAT handling
: Copy data into/out of the DAT module.
Instrument version
: Device internal data e.g. serial
number can be queried.
ChemoClean
(only, if the complete ChemoClean
function is activated): Inspecting programmes, inputs,
mechanics.
Factory function
: Reset counter, write access