Commissioning via operating menu
Micropilot FMR51, FMR52 HART
72
Hauser
11.7 Configuration management
After commissioning, you can save the current device configuration, copy it to another
measuring point or restore the previous device configuration. You can do so using the
Configuration management
parameter and its options.
Navigation path in the operating menu
Setup → Advanced setup → Configuration backup display → Configuration management
Meaning of the options
•
Cancel
No action is executed and the user exits the parameter.
•
Execute backup
A backup copy of the current device configuration in the HistoROM (built-in in the
device) is saved to the display module of the device. The backup copy comprises the
transmitter and sensor data of the device.
•
Restore
The last backup copy of the device configuration is copied from the display module to the
HistoROM of the device. The backup copy comprises the transmitter and sensor data of
the device.
•
Duplicate
The transmitter configuration is duplicated to another device using the transmitter
display module. The following parameters, which characterize the individual measuring
point are
not
included in the transmitted configuration:
– HART date code
– HART short tag
– HART message
– HART descriptor
– HART address
– Device tag
– Medium type
•
Compare
The device configuration saved in the display module is compared to the current device
configuration of the HistoROM. The result of this comparison is displayed in the
Comparison result
parameter.
•
Clear backup data
The backup copy of the device configuration is deleted from the display module of the
device.
While this action is in progress, the configuration cannot be edited via the local
display and a message on the processing status appears on the display.
If an existing backup is restored to a different device using the
Restore
option, it may
occur that some device functionalities are no longer available. In some cases even a
device reset → 157 will not restore the original status.
In order to transmit a configuration to a different device, the
Duplicate
option should
always be used.