Micropilot FMR51, FMR52 HART
Operating menu
Hauser
139
"Current output 1 to 2" submenu
The
Current output 2
submenu (→ 139) is only available for devices with two
current outputs.
Navigation
Setup → Advanced setup → Curr.output 1 to 2
Assign current output 1 to 2
Navigation
Setup → Advanced setup → Curr.output 1 to 2 → Assign curr.
Description
Select process variable for current output.
Selection
• Level linearized
• Distance
• Electronic temperature
• Relative echo amplitude
• Analog output adv. diagnostics 1
• Analog output adv. diagnostics 2
• Area of incoupling
Factory setting
• Current output 1: Level linearized
• Current output 2
: Relative echo amplitude
Additional information
Definition of the current range for the process variables
Process variable
4 mA value
20 mA value
Level linearized
0 %
or the associated
linearized value
100 %
or the associated linearized value
Distance
0 (i.e. level is at the reference
point)
0 %)
Electronic temperature
–50 °C (–58 °F)
100 °C (212 °F)
Relative echo amplitude
0 dB
150 dB
Analog output adv.
diagnostics 1/2
depending on the parametrization of the Advanced Diagnostics
Area of incoupling
0
100
1)
the 0% level is defined by
Empty calibration
2)
The 100% level is defined by
Full calibration
It may be necessary to adjust the 4mA and 20mA values to the application (especially
in the case of the
Analog output adv. diagnostics 1/2
and
Area of incoupling
options).
This can be done by the following parameters:
• Expert → Output → Curr.output 1 to 2 → Turn down
• Expert → Output → Curr.output 1 to 2 → 4 mA value
• Expert → Output → Curr.output 1 to 2 → 20 mA value
8)
only for devices with two current outputs