Electrical connection
Micropilot FMR51, FMR52 HART
42
Hauser
7.1.4
Supply voltage
2-wire, 4-20mA HART, passive
"Power Supply, Output"
"Approval"
Terminal voltage U at the
device
Maximum load R, depending on the supply voltage U
0
at the
supply unit
A:
2-wire; 4-20mA HART • Non-Ex
• Ex nA
• Ex ic
• CSA GP
10.4 to 35 V
R [ ]
?
U0 [V]
10
10.4
21.4
20
30
35
0
500
A0017140
Ex ia / IS
10.4 to 30 V
• Ex d(ia) / XP
• Ex ic(ia)
• Ex nA(ia)
• Ex ta / DIP
12 to 35 V
R [ ]
W
U0 [V]
10
12
23
20
30
35
0
500
A0019136
Ex ia + Ex d(ia) / IS + XP 12 to 30 V
1)
Feature 020 of the product structure
2)
Feature 010 of the product structure
3)
For ambient temperatures T
a
≤ -20 °C (-4 °F) a minimum voltage of 15 V is required for the sartup of the device at the MIN error current (3,6
mA). The startup current can be parametrized. If the device is operated with a fixed current I ≥ 5,5 mA (HART multidrop mode), a voltage of U ≥
10,4 V is sufficient throughout the entire range of ambient temperatures.
4)
For ambient temperatures T
a
≤ -20 °C (-4 °F) a minimum voltage of 16 V is required for the startup of the device at the MIN error current (3.6
mA).
"Power Supply, Output"
"Approval"
Terminal voltage U
at the device
Maximum load R, depending on the supply voltage U
0
at the
supply unit
B:
2-wire; 4-20 mA HART,
switch output
• Non-Ex
• Ex nA
• Ex nA(ia)
• Ex ic
• Ex ic(ia)
• Ex d(ia) / XP
• Ex ta / DIP
• CSA GP
12 to 35 V
R [ ]
W
U0 [V]
10
12
23
20
30
35
0
500
A0019136
• Ex ia / IS
• Ex ia + Ex d(ia) / IS + XP
12 to 30 V
1)
Feature 020 of the product structure
2)
Feature 010 of the product structure
3)
For ambient temperatures T
a
≤ -30 °C (-22 °F) a minimum voltage of 16 V is required for the startup of the device at the MIN error current (3.6
mA).
Summary of Contents for Micropilot FMR50