Description of device parameters
Deltabar FMD71, FMD72
24
Pos. Zero Adjust (007)
Navigation
Expert → Measurement → Basic Setup → Pos. Zero Adjust (007)
Write permission
Operators/Service engineers/Expert
Description
Position adjustment – the pressure difference between zero (set point) and the measured
pressure need not be known.
Example
• Measured value = 2.2 mbar (0.033 psi)
• You correct the measured value via the "Pos. Zero Adjust" parameter with the "Confirm"
option. This means that you are assigning the value 0.0 to the pressure present.
• Measured value (after pos. zero adjust) = 0.0 mbar
• The current value is also corrected.
Options
• Confirm
• Cancel
Factory setting
Cancel
Calib. Offset (008)
Navigation
Expert → Measurement → Basic Setup → Calib. Offset (008)
Write permission
Service engineers/Expert
Description
Position adjustment – the pressure difference between the set point and the measured
pressure must be known.
Example
• Measured value = 982.2 mbar (14.73 psi)
• You correct the measured value with the value entered (e.g. 2.2 mbar (0.033 psi)) via
the "Calib. Offset" parameter. This means that you are assigning the value 980.0 (14.7
psi) to the pressure present.
• Measured value (after pos. zero adjust) = 980.0 mbar (14.7 psi)
• The current value is also corrected.
Factory setting
0.0
Empty Calib. (011/28)
Navigation
Expert → Measurement → Level → Empty Calib. (011/028)
Write permission
Operators/Service engineers/Expert
Description
Enter the output value for the lower calibration point (vessel empty). The unit defined in
"Unit Before Lin." must be used.