Operation with an operating menu
Deltabar FMD71, FMD72
12
www.de.endress.com → Search: FieldCare → FieldCare → Technical data.
FieldCare supports the following functions:
• Configuration of transmitters in online/offline mode
• Loading and saving device data (upload/download)
• Documentation of the measuring point
2.6
Direct access to parameters
The parameters can only be accessed directly via the "Expert" user role.
Direct access (119)
Navigation
Expert → Direct access
Read permission
Operators/Service engineers/Expert
Write permission
Operators/Service engineers/Expert
Description
Enter the direct access code to go directly to a parameter.
Options
Enter the desired parameter code.
Factory setting
0
Note
For direct access, it is not necessary to enter leading zeros.
2.7
Locking/unlocking operation
Once you have entered all the parameters, you can lock your entries against unauthorized
and undesired access.
You have the following options for locking/unlocking operation:
• Via the DIP switch on the electronic insert, local at the device.
• Via the local display (optional)
• Via communication e.g. FieldCare and HART handheld terminal.
The
symbol on the local display indicates that operation is locked. Parameters which
refer to how the display appears, e.g. "Language" and "Display contrast", can still be altered.
If operation is locked by means of the DIP switch, you can only unlock operation again
by means of the DIP switch. If operation is locked by means of local display or remote
operation e.g. FieldCare, you can unlock operation either using the local display or
remote operation.
The "Operator Code" parameter is used to lock and unlock the device.
The parameters can only be accessed directly via the "Expert" user role.
Operator Code (021)
Navigation
Setup → Advanced Setup → Operator Code
Read permission
Operators/Service engineers/Expert
Summary of Contents for Deltabar FMD71